The HM-JACKarc analyser is a compact bench top system that uses Integrated Sphere Latex Turbidimetry to measure faecal haemoglobin concentration using a highly sensitive latex reagent coated with polyclonal antibodies to human haemoglobin.
The instrument has advanced innovative software with touch screen technology, providing a very easy to use interface. There is minimal depth to the programme menus, ensuring quick access to operating screens.
The system can be utilised by all grades of staff with minimal training. Different levels of user access authority provide security, so that only advanced users can amend settings etc.
HM-JACKarc has been designed to provide rapid results with maximum convenience for the operator.
The ability to load up to 80 samples at any one time provides walk away capability and enhanced productivity. Additional consumables and samples (up to a total maximum of 80) can be added at any time during the testing process.
HM-JACKarc has a time to first result of just 5.6 minutes with additional results every 18 seconds, generating up to 200 results per hour.
The technology and software ensures that the system can be used in either screening or symptomatic assessment applications, since cut-off concentrations can be selected depending upon requirements.
Key features of HM-JACKarc FIT System
- Fully automated, compact and light bench top unit with touchscreen interface
- Uses Integrated Sphere Latex Turbidimetry to measure faecal haemoglobin concentration
- Sensitivity: 7 ng/mL
- User definable cut-off to differentiate between Symptomatic or Screening use
- No prozone effect up to 200,000 ng/mL
- High speed performance: 200 samples/hour
- Time to first result 5.6 minutes
- Additional results every 18 seconds
- Factory set Master curve with local 2 point recalibration
- Calibrates system to local conditions and reagent combinations
- Calibration weekly
- System can store 2 calibration curves
- Calibration of different lots of latex
- Same lots with different calibrators
HM-JACKarc FIT System Specification
System Specifications: | |
---|---|
Analysis Technology | Latex Particle Agglutination |
Measurement Principles | Integrated Spherical Nephelometry (Turbidity) |
Time to First Result | 5.6 minutes |
Sample Capacity | 80 samples (8 x 10 position racks) |
Calibration (With local correction using a 2 level calibrator) | Factory set 7 point curve |
Data Output | RS232 serial port |
Integrated Printer | 32- figure thermal printer |
Power Supply | AC220V +/- 10V |
Power Consumption | 300 VA |
Dimensions | H 500mm x W 600mm x D 610mm |
Weight | 56 Kg |